Inheco OVT - ODTC® Verification Tool

Check your On Deck Thermal Cycler is working perfectly

Verify your ODTC® performance on a regular basis or before you process critical samples or run your assay.
The ODTC® Verification Tool is purpose-built to measure ODTC® performance in real operating situations. It’s designed for end-users or maintenance staff to check Installation Qualification (IQ) and Operational Qualification (OQ) as well as for regular maintenance.

EASY
The OVT is easy to use, with step-by-step guidance included to set it up. It uses a USB connection and an automated test routine. In just 25 minutes, the OVT can assess the ODTC®’s performance. The system software automatically calculates heating & cooling rates, precision and accuracy in the mount and lid, and temperature uniformity during the plateau phase. System information and test results, plus your own personalized information, are delivered in an audit-quality PDF certificate.

ACCURATE
We ensure near -real operating conditions - the OVT fits perfectly when the ODTC® lid is closed, and thanks to its unique flat measuring head. The OVT checks heating and cooling rates across the Vapor Chamber Mount during a predefined test routine that compares accurately to real PCR runs. Sensors also verify the lid temperature.

MATCH to your ODTC®
There are two versions of the OVT available, OVT96 for verification of all ODTC® 96 well versions, and OVT384 for verification of all ODTC® 384 well versions.

RELIABLE and SUPPORTED
Your OVT is delivered with an initial temperature calibration from an accredited calibration laboratory and must be recalibrated every year. Contact Inheco tech support to arrange calibration.

Every Inheco accessory is designed, engineered and made in Germany and undergoes rigorous quality control before delivery. It comes with a global two-year warranty and full customer service including remote support, diagnostic and servicing options that focusses on minimizing operational downtime.

Get a Quote
 

 
 

Technical Specifications OVT

General Specifications - OVT Measurement Head 1)
Dimensions (WxDxH) 96 mm x 126 mm x 9.4 mm (with sensor pins H: 20.5 mm (96) and 14.7 mm (384))
Weight 100 g
Protection Class IP 202)
Verfication of ODTC® 96 and ODTC® 384
Number of Sensing Elements Mount verification 5 Sensing Elements. Lid verification 2 Sensing Elements

1) OVT Measurement Head shall only be connected to the OVT Data Logger.
2) not protected against water.
 
Thermal Specifications - OVT Measurement Head
Temperature range (absolute) Mount +4°C to +99°C [+39.2°F to +210°F]
Lid +30°C to +120°C [+86F to +248°F]
Temperature accuracy (absolute) Mount steady state: ±0.10 K at +50°C and +90°C [+122°F and +194°F ]
Lid steady state: ±0.20 K at +50°C and +90°C [+122°F and +194°F ]
Temperature uniformity (relative) Mount steady state: ±0.05 K at +50°C and +90°C [+122°F and +194°F]
 
Specifications - OVT Data Logger
Dimensions (WxDxH) 116 mm x 216 mm x 64 mm
Interface to PC USB 2.0 compatible
noise level No noise
Weight 700 g
DC Input 5V
Protection Class IP30
 
Environmental Conditions - OVT Data Logger
Tolerable relative humidity:
Operation RH 30% to 80% ( non condensing ) 3)
Transportation and storage RH 10% to 95% ( non condensing ) 3)
Temperature:
Operation +18°C to + 28°C [+64.4°F and +82.4°F] altitude 0-2,000m
Transportation and storage -10°C to +60°C [+14°F and +140°F] altitude 0-11,000m

3) Condensate can prevent the OVT® from operating properly and can damage the OVT®
Inheco
Industrial Heating & Cooling GmbH
Fraunhoferstrasse 11
82152 Martinsried / Germany
 
Inheco Sales
Phone: +49 89 899593-120
Email: sales@inheco.com
 
Inheco Technical Support
Phone: +49 89 899593-121
Email: techhotline@inheco.com
 
Inheco People & Culture
Phone: +49 89 899593-421
Email: jobs@inheco.com